The AVOX 801238 Series™ Inhalator incorporates all of the components of the AVOX Pressur Vak II™ with the exception of the face seal and manifold assembly. The AVOX 801238 Series™ Inhalator with overboard discharge provides control for the administration of oxygen or gas mixtures without affecting the chamber environment by exhausting the exhaled gases out of the chamber.
In most instances, the vacuum system is the differential pressure between the inside and outside of the chamber. Breathing gas is reduced to approximately 65 psig to 125 psig above ambient pressure. The low pressure gas flows through a hose to the demand regulator. The exhalation is removed through the vacuum regulator and exhaust hose. For a pressure differential in excess of 60 FSW, a second vacuum regulator must be installed between the hose and chamber wall. The Tescom 26-2912-282A Regulator will suffice. The secondary regulator will limit the pressure differential across the vacuum regulator on the mask.
The AVOX 801238-01 Series™ Inhalator comes complete with Demand and Exhaust Dual Hose Assembly with Quick Disconnects, Demand Regulator Assembly and Exhaust Regulator Assembly.
Warning: The AVOX 801238 Series™ Inhalator with overboard discharge is used for surface decompression of persons in a recompression chamber and is not for use underwater. The Mask may also be used in other applications where overboard discharge of the user’s exhalation is desired.